Calibration Substrate Selection Guide | |||||
---|---|---|---|---|---|
Calibration Substrate |
Pad Size (microns) |
Calibration Types Supported |
Footprint |
Pitch Range Recommended (microns) |
Pitch Range Acceptable (microns) |
CS-5 | 50 X 50 |
SOLT, LRL, LRM |
GSG | 75 - 250 | 75 -250 |
CS-9 | 100 X 100 |
SOLT, LRL, LRM |
GSG | 250 - 600 | 150 - 600 |
CS-10 | 150 X 150 | SOLT, LRM | GSG | 600 - 1250 | 225 - 1250 |
CS-18
|
300 X 300 | SOLT, LRM | GSG | 1250 - 2540 | 500 - 2540 |
FIGURE 2 |
CS-8 |
50 X 50 100 X 100 150 X 150 |
SOLT, LRM | GS, SG | 50 - 200 | 50 -300 |
---|---|---|---|---|---|
CS-14 | 100 X 100 | SOLT, LRM | GS, SG | 200 - 400 | 150 - 600 |
CS-11 | 150 X 150 | SOLT, LRM | GS, SG | 400 - 1250 | 175 - 1250 |
CS-17 | 300 X 300 | SOLT, LRM | GS, SG | 750 - 2540 | 450 - 2540 |
FIGURE 3 |
Special Calibration Substrate Designed For Use Above 110 GHz | |||||
---|---|---|---|---|---|
Calibration Substrate |
Pad Size (microns) |
Calibration Types Supported |
Footprint |
Pitch Range Recommended (microns) |
Pitch Range Acceptable (microns) |
CS-15 | 25 X 25 |
SOLT, LRL, LRM |
GSG |
40 - 150 (SOLT) 30 - 150 (LRL) |
40 - 150 |
FIGURE 4 |